Biomedical Engineering / Blood Rheology Casson Model
FREE TO EXPLORE
LABORATORY 27 · BIOMEDICAL ENGINEERING

Blood Rheology Casson Model lab report

A worked blood rheology casson model lab report with sample observation tables, regression plots, calculations, and error analysis.

WORKED EXAMPLE · BENCHMARK OBSERVATIONS

A complete example laboratory report.

Simulated demonstration data.

This report illustrates formal biomedical engineering data reduction, transducer calibration tares, and physiological parameter estimation. Replace its observations with your own experimental telemetry when preparing a biomedical engineering laboratory report submission.

Student: __________________Student ID: __________________Course: __________________Date: __________________Partner: __________________Instructor: __________________

1. Title and aim

Non-Newtonian Blood Rheology & Casson Fluid Model Cone-and-Plate Viscometry. Determine the Casson asymptotic plastic viscosity μ_c and yield stress τ_y of whole human blood across low-to-medium shear rates using cone-and-plate rheometry.

2. Abstract

Six controlled biomedical engineering observations were evaluated across standardized settings Shear Rate γ̇ = 16.0 s⁻¹ to Shear Rate γ̇ = 196.0 s⁻¹ for Non-Newtonian Blood Rheology & Casson Fluid Model Cone-and-Plate Viscometry. An ordinary least-squares linear regression yielded an empirical Casson Plastic Viscosity Root √μ_c of 0.0620 Pa⁰·⁵·s⁰·⁵ (difference from theoretical reference = 0.0000%). The empirical telemetry conclusively validates the fundamental biomedical engineering formulation with R² = 1.000000. The investigation demonstrates stable physiological response, transducer linearity, and reproducible telemetry across all clinical test conditions with negligible residual variance. Data acquired through the virtual biomedical workbench exhibit strict adherence to established biopotential, constitutive biomechanical, hemodynamic, or diagnostic imaging governing equations, confirming the dimensional and similitude assertions of the curriculum.

3. Apparatus and fixed settings

4. Theory and graph choice

Human blood is not a simple homogeneous Newtonian fluid; it is a dense suspension of cellular elements (erythrocytes/RBCs, leukocytes, and platelets) suspended in an aqueous electrolyte-protein matrix (blood plasma). Erythrocytes comprise approximately 40% to 45% of total blood volume (normal hematocrit Hct). At low shear rates (γ̇ < 50 s⁻¹), red blood cells aggregate face-to-face into reversible, rouleaux structures and three-dimensional cellular networks stabilized by plasma fibrinogen and globulins. This cellular networking creates shear-thinning (pseudoplastic) non-Newtonian flow behavior and a finite yield stress τ_y below which blood resists flow. At elevated shear rates, rouleaux disperse into individual deformable biconcave discs that elongate and align with streamlines, causing apparent viscosity to asymptote toward a constant high-shear baseline. To mathematically characterize this non-Newtonian behavior, the Casson fluid model is the universally adopted standard: √τ = √μ_c · √γ̇ + √τ_y, where τ is shear stress, γ̇ is shear rate, μ_c is the Casson asymptotic plastic viscosity (Pa·s), and τ_y is the blood yield stress (Pa). In a cone-and-plate rotational rheometer, the shear rate is geometrically uniform throughout the entire wedge gap, providing ideal conditions to evaluate blood rheology. In this laboratory investigation, anti-coagulated whole blood at 37.0 °C is tested across shear rates from 16 to 196 s⁻¹ to determine Casson parameters.

√τ = √μ_c · √γ̇ + √τ_y = 0.0620 · √γ̇ + 0.1800

The governing physiological relationship was linearized into the standard regression coordinate form y = mx + b. Here, the independent coordinate x represents Square root of shear rate √γ̇ (s⁻⁰·⁵), and the dependent coordinate y represents Square root of shear stress √τ (Pa⁰·⁵). The gradient m directly determines Casson Plastic Viscosity Root √μ_c, while the intercept b accounts for baseline null offsets, pre-stress tares, or reference thresholds. Linear transformation allows robust parameter estimation via ordinary least squares without non-linear solver instability or convergence traps. Plotting transformed observation pairs displays sensor sensitivities, elastic moduli, hemodynamic resistances, or acoustic attenuation coefficients directly as linear slopes. This linear graphical formulation provides an immediate visual check of homoscedasticity, linearity, and potential sensor saturation, biological yield, or non-linear compliance.

5. Method

  1. Mix EDTA-anticoagulated whole blood on roller mixer for 10 minutes; verify standard hematocrit (Hct = 45 ± 1%) via microhematocrit centrifuge.
  2. Set rheometer Peltier bottom plate to exactly 37.0 °C and allow 15 minutes for thermal equilibrium.
  3. Zero the normal force and calibrate the mechanical contact gap; pipette exactly 500 μL of blood onto the plate center and lower cone to truncation height (50 μm).
  4. Engage the humidified solvent trap and apply a 30-second preshear at 300 s⁻¹ to completely disperse pre-existing rouleaux aggregates.
  5. Step through shear rates γ̇ = 16, 36, 64, 100, 144, and 196 s⁻¹, holding each step for 15 seconds to achieve steady torque before recording shear stress τ.
  6. Compute √γ̇ and √τ, plot √τ versus √γ̇, and extract Casson asymptotic viscosity root √μ_c from the slope and yield stress root √τ_y from the y-intercept.

6. Observations and calculation table

All biomedical observations use strictly controlled 64-bit IEEE 754 precision buffers. Residuals are computed relative to the ordinary least-squares regression line.

Table 1. Six simulated observations; x and y represent transformed biomedical metrics.
No.Fluid Shear Rate γ̇ (s⁻¹)Measured Shear Stress τ (Pa)Square root of shear rate √γ̇ (s⁻⁰·⁵)Square root of shear stress √τ (Pa⁰·⁵)Fitted yResidual
116.00.18324.00000.42800.4280-0.0000
236.00.30476.00000.55200.55200.0000
364.00.4578.00000.67600.67600.0000
4100.00.6410.00000.80000.80000.0000
5144.00.853812.00000.92400.92400.0000
6196.01.098314.00001.04801.04800.0000

Download the sample observations (CSV) ↓

7. Worked calculation for one observation

For input Shear Rate γ̇ = 16.0 s⁻¹: Measured Measured Shear Stress τ = 0.1832 Pa. Transformed biomedical coordinates: independent x = 4.0000, dependent y = 0.4280.

8. Graph, gradient and final calculation

0.332.800.545.900.749.000.9412.101.1415.20Square root of shear rate √γ̇ (s⁻⁰·⁵)Square root of shear stress √τ (Pa⁰·⁵)

Circular markers represent empirical biomedical observations. The solid line represents the ordinary least-squares linear fit.

x̄ = 9.0000; ȳ = 0.7380
Σ(x − x̄)² = 70.0000
Σ(x − x̄)(y − ȳ) = 4.3400
m = Σ(x − x̄)(y − ȳ) / Σ(x − x̄)² = 0.06200000
b = ȳ − mx̄ = 0.18000000
y = 0.0620 · x + 0.1800 ⟹ Target Casson Plastic Viscosity Root √μ_c = 0.0620 Pa⁰·⁵·s⁰·⁵ (theoretical = 0.0620)

9. Uncertainty and comparison

Percentage difference = |estimate − theoretical value| / |theoretical value| × 100 = 0.0000%.

For n = 6 observations, Σresidual² = 3.081e-33; s² = Σresidual²/(n − 2). Gradient standard error = √[s²/Σ(x − x̄)²] = 3.317e-18. Intercept standard error = √[s²(1/n + x̄²/Σ(x − x̄)²)] = 3.193e-17.

Propagating the fit parameters yields a fit standard error of 3.317e-18 Pa⁰·⁵·s⁰·⁵, with an ideal coefficient of determination R² = 1.000000.

In biomedical engineering experimentation, clinical telemetry, and medical device verification, empirical uncertainties originate from physiological biological variability, electrode-skin electrochemical polarization, transducer thermal drift, amplifier analog-to-digital quantization noise (64-bit IEEE 754 precision), and fluid boundary layer fluctuations. The ordinary least-squares regression across n = 6 standardized test intervals yields unbiased minimum-variance estimators for both slope and intercept. Because the underlying bioinstrumentation, continuum biomechanics, hemodynamics, and imaging physics adhere strictly to conservation of mass, momentum, charge, and electromagnetic principles, residual dispersions are tightly bounded, yielding an ideal coefficient of determination R² ≥ 0.999999 and validating the theoretical formulation. Furthermore, systematic errors—including amplifier common-mode rejection offsets, sensor cable movement artifacts, hydrostatic fluid leveling offsets, and specimen hydration boundaries—are corrected according to medical instrumentation protocols (Webster, Bronzino, and ISO standards). Standard deviations are propagated using first-order multivariate Taylor expansions (delta method) across transformed physiological coordinates. Residual diagnostics confirm homoscedasticity across the entire operational clinical window without systematic non-linear bias.

10. Result and discussion

Casson Plastic Viscosity Root √μ_c = 0.0620 Pa⁰·⁵·s⁰·⁵
Theoretical reference value: 0.0620 Pa⁰·⁵·s⁰·⁵

The empirical laboratory observations provide rigorous quantitative verification of Non-Newtonian Blood Rheology & Casson Fluid Model Cone-and-Plate Viscometry. In biomedical engineering, medical instrumentation, biomechanics, and artificial organs, physical systems are governed by Maxwell equations, Navier-Stokes fluid mechanics, Hookean continuum elasticity, and biphasic mixture theories. Across all six observation trials, the empirical slope m = 0.0620 confirms the theoretical model with near-zero residual variance. Furthermore, comparison with established medical engineering benchmarks (Webster, Bronzino, Fung, and Winter) underscores vital clinical design principles including patient electrical safety, biocompatibility, transducer signal-to-noise ratio, fatigue longevity, and hemodynamic shear preservation. Residual analysis confirms homoscedastic dispersion without systematic non-linear patterning, validating the underlying model assumptions. Confidence intervals calculated via asymptotic standard errors align with theoretical quantiles, indicating that finite-sample distortions remain negligible under controlled test conditions. The experimental protocol confirms the robustness of the methodology across varying physiological operating states.

11. Precautions and limitations

The Casson model is an empirical continuum approximation; it applies across 10 to 300 s⁻¹ but does not capture zero-shear viscoelasticity. Evaluated on anti-coagulated human whole blood (Hct = 45%, EDTA treated) at 37.0 °C in a titanium cone-and-plate rheometer (cone angle θ = 1.0°, diameter 60 mm). Standard shear rates γ̇ = 16, 36, 64, 100, 144, 196 s⁻¹ correspond to √γ̇ = 4, 6, 8, 10, 12, 14 s⁻⁰·⁵. Casson slope √μ_c = 0.0620 Pa⁰·⁵·s⁰·⁵ (asymptotic viscosity μ_c = 0.00384 Pa·s = 3.84 cP) with yield root √τ_y = 0.1800 Pa⁰·⁵ (yield stress τ_y = 0.0324 Pa).

12. Conclusion

This laboratory investigation quantitatively verified the mathematical formulation and physical behavior of Non-Newtonian Blood Rheology & Casson Fluid Model Cone-and-Plate Viscometry. Through rigorous tracking across six observation points, we confirmed an empirical Casson Plastic Viscosity Root √μ_c of 0.0620 Pa⁰·⁵·s⁰·⁵ matching the theoretical reference 0.0620 Pa⁰·⁵·s⁰·⁵ within 0.0000% error. The investigation demonstrates why this biomedical engineering method serves as an indispensable cornerstone in modern clinical patient monitoring, implant design, and diagnostic systems. The verified scaling laws and parameter estimates demonstrate the practical efficacy of physiological modeling and biomedical engineering principles.