Biomedical Engineering / Instrumentation Amplifier CMRR
FREE TO EXPLORE
LABORATORY 05 · BIOMEDICAL ENGINEERING

Instrumentation Amplifier CMRR lab report

A worked instrumentation amplifier cmrr lab report with sample observation tables, regression plots, calculations, and error analysis.

WORKED EXAMPLE · BENCHMARK OBSERVATIONS

A complete example laboratory report.

Simulated demonstration data.

This report illustrates formal biomedical engineering data reduction, transducer calibration tares, and physiological parameter estimation. Replace its observations with your own experimental telemetry when preparing a biomedical engineering laboratory report submission.

Student: __________________Student ID: __________________Course: __________________Date: __________________Partner: __________________Instructor: __________________

1. Title and aim

Three-Op-Amp Instrumentation Amplifier Common-Mode Rejection Ratio (CMRR). Measure differential gain and common-mode gain of a classic three-op-amp biomedical instrumentation amplifier to calculate CMRR across common-mode interference voltages.

2. Abstract

Six controlled biomedical engineering observations were evaluated across standardized settings V_cm = 0.5 V to V_cm = 3.0 V for Three-Op-Amp Instrumentation Amplifier Common-Mode Rejection Ratio (CMRR). An ordinary least-squares linear regression yielded an empirical Common-Mode Voltage Gain A_cm of 0.0480 mV/V (difference from theoretical reference = 0.0000%). The empirical telemetry conclusively validates the fundamental biomedical engineering formulation with R² = 1.000000. The investigation demonstrates stable physiological response, transducer linearity, and reproducible telemetry across all clinical test conditions with negligible residual variance. Data acquired through the virtual biomedical workbench exhibit strict adherence to established biopotential, constitutive biomechanical, hemodynamic, or diagnostic imaging governing equations, confirming the dimensional and similitude assertions of the curriculum.

3. Apparatus and fixed settings

4. Theory and graph choice

In clinical biopotential recording (ECG, EEG, EMG), the target physiological signal is exceedingly small—typically 1.0 mV for ECG and under 50 μV for EEG. Simultaneously, the human body acts as an antenna that capacitively couples to 50/60 Hz powerline electric fields, inducing large common-mode interference voltages (often 1.0 V to 3.0 V peak-to-peak, thousands of times larger than the bioelectric signal). Extracting microvolt biopotentials in the presence of massive common-mode noise requires an analog front-end with an extraordinary Common-Mode Rejection Ratio (CMRR > 90 to 110 dB). The classic three-operational-amplifier instrumentation amplifier (INA) achieves this through a buffered input stage with gain R_G: differential gain is set by A_d = (1 + 2·R_1 / R_G) · (R_3 / R_2), while ideal common-mode gain is A_cm = 0. In physical hardware, small resistor mismatches (tolerance error ΔR/R) in the second-stage difference amplifier allow a fraction of common-mode input voltage V_cm to leak into the output: V_out,cm = A_cm · V_cm. The CMRR in decibels is defined as CMRR = 20 · log10(A_d / A_cm). In this laboratory experiment, a precision biomedical instrumentation amplifier (configured for differential gain A_d = 1000 V/V) is tested across regulated common-mode test voltages to quantify A_cm, verify linear leakage, and evaluate front-end CMRR.

V_out,cm = A_cm · V_cm = 0.0480 · V_cm

The governing physiological relationship was linearized into the standard regression coordinate form y = mx + b. Here, the independent coordinate x represents Common-mode input V_cm (V), and the dependent coordinate y represents Common-mode output V_out,cm (mV). The gradient m directly determines Common-Mode Voltage Gain A_cm, while the intercept b accounts for baseline null offsets, pre-stress tares, or reference thresholds. Linear transformation allows robust parameter estimation via ordinary least squares without non-linear solver instability or convergence traps. Plotting transformed observation pairs displays sensor sensitivities, elastic moduli, hemodynamic resistances, or acoustic attenuation coefficients directly as linear slopes. This linear graphical formulation provides an immediate visual check of homoscedasticity, linearity, and potential sensor saturation, biological yield, or non-linear compliance.

5. Method

  1. Apply a 1.000 mV peak-to-peak 10 Hz pure differential input signal; verify that differential output reads exactly 1.000 V peak-to-peak, confirming A_d = 1000 V/V.
  2. Tie both differential non-inverting and inverting inputs together to apply identical common-mode voltage V_cm simultaneously.
  3. Apply a 60 Hz sinusoidal common-mode voltage of 0.500 V RMS, monitoring amplifier rails to ensure linear range.
  4. Measure the residual 60 Hz output voltage V_out,cm using the precision true-RMS microvoltmeter over a 30-second integration window.
  5. Increment common-mode voltage V_cm through 1.0, 1.5, 2.0, 2.5, and 3.0 V RMS, recording V_out,cm at each stage.
  6. Plot V_out,cm against V_cm to determine common-mode gain A_cm = slope / 1000 (in V/V), then calculate CMRR = 20·log10(A_d / A_cm).

6. Observations and calculation table

All biomedical observations use strictly controlled 64-bit IEEE 754 precision buffers. Residuals are computed relative to the ordinary least-squares regression line.

Table 1. Six simulated observations; x and y represent transformed biomedical metrics.
No.Common-Mode Input Voltage V_cm (V)Common-Mode Output Voltage V_out,cm (mV)Common-mode input V_cm (V)Common-mode output V_out,cm (mV)Fitted yResidual
10.50.0240.50000.02400.02400.0000
21.00.0481.00000.04800.04800.0000
31.50.0721.50000.07200.0720-0.0000
42.00.0962.00000.09600.09600.0000
52.50.122.50000.12000.12000.0000
63.00.1443.00000.14400.1440-0.0000

Download the sample observations (CSV) ↓

7. Worked calculation for one observation

For input V_cm = 0.5 V: Measured Common-Mode Output Voltage V_out,cm = 0.0240 mV. Transformed biomedical coordinates: independent x = 0.5000, dependent y = 0.0240.

8. Graph, gradient and final calculation

0.010.200.040.970.081.750.122.520.163.30Common-mode input V_cm (V)Common-mode output V_out,cm (mV)

Circular markers represent empirical biomedical observations. The solid line represents the ordinary least-squares linear fit.

x̄ = 1.7500; ȳ = 0.0840
Σ(x − x̄)² = 4.3750
Σ(x − x̄)(y − ȳ) = 0.2100
m = Σ(x − x̄)(y − ȳ) / Σ(x − x̄)² = 0.04800000
b = ȳ − mx̄ = 0.00000000
y = 0.0480 · x + 0.0000 ⟹ Target Common-Mode Voltage Gain A_cm = 0.0480 mV/V (theoretical = 0.0480)

9. Uncertainty and comparison

Percentage difference = |estimate − theoretical value| / |theoretical value| × 100 = 0.0000%.

For n = 6 observations, Σresidual² = 9.630e-34; s² = Σresidual²/(n − 2). Gradient standard error = √[s²/Σ(x − x̄)²] = 7.418e-18. Intercept standard error = √[s²(1/n + x̄²/Σ(x − x̄)²)] = 1.444e-17.

Propagating the fit parameters yields a fit standard error of 7.418e-18 mV/V, with an ideal coefficient of determination R² = 1.000000.

In biomedical engineering experimentation, clinical telemetry, and medical device verification, empirical uncertainties originate from physiological biological variability, electrode-skin electrochemical polarization, transducer thermal drift, amplifier analog-to-digital quantization noise (64-bit IEEE 754 precision), and fluid boundary layer fluctuations. The ordinary least-squares regression across n = 6 standardized test intervals yields unbiased minimum-variance estimators for both slope and intercept. Because the underlying bioinstrumentation, continuum biomechanics, hemodynamics, and imaging physics adhere strictly to conservation of mass, momentum, charge, and electromagnetic principles, residual dispersions are tightly bounded, yielding an ideal coefficient of determination R² ≥ 0.999999 and validating the theoretical formulation. Furthermore, systematic errors—including amplifier common-mode rejection offsets, sensor cable movement artifacts, hydrostatic fluid leveling offsets, and specimen hydration boundaries—are corrected according to medical instrumentation protocols (Webster, Bronzino, and ISO standards). Standard deviations are propagated using first-order multivariate Taylor expansions (delta method) across transformed physiological coordinates. Residual diagnostics confirm homoscedasticity across the entire operational clinical window without systematic non-linear bias.

10. Result and discussion

Common-Mode Voltage Gain A_cm = 0.0480 mV/V
Theoretical reference value: 0.0480 mV/V

The empirical laboratory observations provide rigorous quantitative verification of Three-Op-Amp Instrumentation Amplifier Common-Mode Rejection Ratio (CMRR). In biomedical engineering, medical instrumentation, biomechanics, and artificial organs, physical systems are governed by Maxwell equations, Navier-Stokes fluid mechanics, Hookean continuum elasticity, and biphasic mixture theories. Across all six observation trials, the empirical slope m = 0.0480 confirms the theoretical model with near-zero residual variance. Furthermore, comparison with established medical engineering benchmarks (Webster, Bronzino, Fung, and Winter) underscores vital clinical design principles including patient electrical safety, biocompatibility, transducer signal-to-noise ratio, fatigue longevity, and hemodynamic shear preservation. Residual analysis confirms homoscedastic dispersion without systematic non-linear patterning, validating the underlying model assumptions. Confidence intervals calculated via asymptotic standard errors align with theoretical quantiles, indicating that finite-sample distortions remain negligible under controlled test conditions. The experimental protocol confirms the robustness of the methodology across varying physiological operating states.

11. Precautions and limitations

The linear leakage model applies within the linear common-mode input range (V_cm ≤ 3.0 V). Evaluated using a low-noise medical instrumentation amplifier (INA118 precision architecture) configured for fixed differential gain A_d = 1000. Common-mode gain A_cm is evaluated across 60 Hz sinusoidal common-mode voltages from 0.5 V to 3.0 V RMS.

12. Conclusion

This laboratory investigation quantitatively verified the mathematical formulation and physical behavior of Three-Op-Amp Instrumentation Amplifier Common-Mode Rejection Ratio (CMRR). Through rigorous tracking across six observation points, we confirmed an empirical Common-Mode Voltage Gain A_cm of 0.0480 mV/V matching the theoretical reference 0.0480 mV/V within 0.0000% error. The investigation demonstrates why this biomedical engineering method serves as an indispensable cornerstone in modern clinical patient monitoring, implant design, and diagnostic systems. The verified scaling laws and parameter estimates demonstrate the practical efficacy of physiological modeling and biomedical engineering principles.