Biomedical Engineering / Windkessel Arterial Compliance
FREE TO EXPLORE
LABORATORY 29 · BIOMEDICAL ENGINEERING

Two-Element Windkessel Lumped-Parameter Arterial Compliance & Diastolic Decay

Determine the arterial Windkessel decay rate constant k_w and evaluate total systemic arterial compliance C_a from the exponential diastolic pressure runoff curve. Explore interactive physiological instrumentation simulation, biomechanics, and biomedical engineering telemetry.

BIOINSTRUMENTATION · BIOSIGNALS · BIOMECHANICS · IMAGING

Your virtual biomedical engineering laboratory.

ARTERIAL PRESSURE RATIO P/P_ES—
PHYSIOLOGICAL & TEST CELL TELEMETRY

Ready to sample with Time t = 300.0 ms.

300.0 ms
System: Mock circulatory flow loop with silicone aorta

Adjust Time t, then execute biomedical trial.

Biomedical engineering telemetry and physiological states are evaluated rigorously using 64-bit IEEE 754 precision. Residuals reflect ordinary least-squares regression across successive observations. Continuous data structures support high-fidelity biological and physiological modeling.

Biomedical assumptions and boundary condition limits

The two-element Windkessel model neglects wave reflection phenomena that create localized systolic pressure amplification in peripheral arteries. Evaluated on a mock cardiovascular circulatory flow loop with a silicone elastic aorta and variable peripheral resistance valve. Systemic parameters: peripheral resistance R_p = 1.00 mmHg·s/mL (133.3 kPa·s/L), end-systolic dicrotic notch pressure P_es = 100 mmHg. Diastolic time constant τ = R_p · C_a = 800 ms, yielding decay rate k_w = 1/800 ms = 0.00125 ms⁻¹ (arterial compliance C_a = 1.067 mL/mmHg). Evaluated across times 100 to 600 ms.

Biomedical apparatus & physiological sensors.

Cardiovascular Mock Circulatory Flow Simulator

Replicates left ventricular stroke volume, tri-leaflet aortic valve, and systemic impedance.

Compliant Silicone Aortic Phantom Chamber

Provides physiological elastic capacitance (compliance C_a ~ 1.0 mL/mmHg) storing systolic volume.

Adjustable Micro-Precision Peripheral Resistance Valve

Calibrates total peripheral resistance R_p to standard physiological value (1.00 mmHg·s/mL).

Catheter-Tipped Solid-State Pressure Transducer (Millar Mikro-Tip)

Records high-fidelity intravascular pressure waveforms at 2000 Hz with zero fluid inertia.

Pneumatic Piston Left Ventricle Pulsatile Driver

Generates 70 mL stroke volumes at 72 beats/min with realistic systolic ejection and diastolic rest times.

Synchronous ECG & Aortic Waveform Telemetry Workstation

Identifies dicrotic notch incisura and isolates diastolic exponential decay segments.

FOLLOW THE METHOD

A reason behind every step.

Determine the arterial Windkessel decay rate constant k_w and evaluate total systemic arterial compliance C_a from the exponential diastolic pressure runoff curve. Standardized parameter settings: 100.0, 200.0, 300.0, 400.0, 500.0, 600.0 ms.

01

Circulatory Loop Priming and Transducer Calibration

Prime flow loop with 40% glycerol-water blood analog (viscosity 3.8 cP at 22 °C), purge air, and calibrate Millar pressure transducer against mercury manometer.

02

Ventricular Hemodynamic Stabilization

Set pneumatic driver to 72 BPM (stroke volume 70 mL, systolic duration 300 ms, diastolic duration 533 ms); verify stable 120/80 mmHg aortic pressure wave.

03

Dicrotic Notch Incisura Identification

Locate the dicrotic notch (closure of aortic valve) marking end-systole (t = 0, P_es = 100.0 mmHg) to define the onset of pure diastolic runoff.

04

Diastolic Pressure Waveform Sampling

Sample decaying arterial pressure P(t) at standardized post-incisura intervals: t = 100, 200, 300, 400, 500, and 600 ms across 10 cardiac cycles.

05

Logarithmic Transformation Calculation

Compute normalized pressure ratio P(t)/P_es and take natural logarithm y = ln(P/P_es) at each time step.

06

Windkessel Decay Regression Analysis

Plot ln(P/P_es) versus decay time t, determine slope -k_w via linear regression, and calculate total arterial compliance C_a = 1 / (R_p · k_w).

YOUR NOTEBOOK · 0 OBSERVATIONS

Turn measurements into biomedical engineering insight.

See a complete example report ↗

Points are your recorded observations. The solid regression line validates Windkessel Arterial Compliance mathematical scaling.

LINEAR REGRESSION FIT · y = mx + b
ln(P/P_es) = -k_w · t = -0.00125 · t
#Diastolic Decay Time t (ms)Arterial Pressure Ratio P/P_es (ratio)Diastolic decay time t (ms)Natural log pressure ratio ln(P/P_es)Remove

No readings yet. Execute a biomedical trial and record it.

Calculate it yourself.

CHECK YOUR UNDERSTANDING

Explain what the biomedical metrics mean.

What mathematical function describes arterial pressure decay during diastole in the two-element Windkessel model?

What happens to aortic compliance C_a as humans age and develop arteriosclerosis?

If the diastolic decay rate constant is k_w = 0.00125 ms⁻¹, what is the arterial time constant τ?

What anatomical event causes the dicrotic notch (incisura) on the aortic pressure tracing?