Chemistry / Surface Tension Stalagmometry
FREE TO EXPLORE
LABORATORY 09 · NATURAL SCIENCES

Surface Tension Stalagmometry lab report

A worked surface tension stalagmometry lab report with gravimetric data tables, linear regression graphs, calculations, uncertainty and conclusions.

WORKED EXAMPLE · QUANTITATIVE OBSERVATIONS

A complete example laboratory report.

Simulated demonstration data.

This report illustrates formal analytical gravimetry, volumetric calibrations, and linear regression analysis. Replace its observations with your own experimental telemetry when preparing a practical submission.

Student: __________________Student ID: __________________Course: __________________Date: __________________Partner: __________________Instructor: __________________

1. Title and aim

Determination of Liquid Surface Tension via Drop-Weight Stalagmometry and Harkins-Jordan Correction. Determine the absolute liquid surface tension of ultra-pure water at 20.0 °C via precision drop-weight stalagmometry, applying Tate's law and Harkins-Jordan hydrodynamic detachment corrections across systematic drop count series, and extracting surface tension using ordinary least-squares regression.

2. Abstract

Six precision drop-weight stalagmometry trials were conducted using ultra-pure deionized water at T = 20.0 ± 0.05 °C (293.15 K) across detached droplet counts from N = 10 to N = 60 drops using a calibrated Traube stalagmometer (external tip radius r = 3.500 mm, Harkins-Jordan correction factor φ = 0.6150). Cumulative detached droplet masses ranged from 1.0040 g to 6.0240 g, collected at a quasi-static detachment rate of 1 drop every 8–10 seconds inside a stoppered weighing bottle. Ordinary least-squares linear regression of cumulative mass against drop count yielded an empirical single-drop mass slope of 0.100400 g/drop, extracting an absolute surface tension of γ = 72.80 mN/m with an intercept of 0.00 g and an ideal coefficient of determination R² = 1.00000, matching the authoritative IAPWS reference standard (72.80 mN/m) with a percentage difference of 0.0000%.

3. Apparatus and fixed settings

4. Theory and graph choice

Surface tension is the fundamental intensive thermodynamic property of a liquid interface that characterizes the reversible work required to expand the interfacial surface area by unit amount: γ = (∂G/∂A)_{T,P,n}. In the interior bulk of a liquid, each molecule experiences isotropic cohesive intermolecular forces pulling uniformly in all directions; at the liquid-vapor boundary, however, molecules lack vapor-phase neighbors and experience a net inward attractive force that contracts the surface to minimal geometric area. In 1864, Thomas Tate formulated the classic law governing droplet detachment from a circular capillary tip: the downward gravitational weight of a detaching drop W = m_drop · g balances the upward capillary tensile force acting along the tip perimeter, W = 2π · r · γ. However, fluid dynamic investigations by Rayleigh, Harkins, and Jordan revealed that Tate's ideal equation assumes complete detachment of the pendent neck; in physical reality, when the liquid neck pinches off via the Rayleigh-Plateau instability, between 35% and 40% of the pendent droplet volume recoils and remains attached to the glass tip. To achieve analytical accuracy, William Draper Harkins and F. E. Jordan (1930) introduced the dimensionless hydrodynamic correction factor φ(r / V^{1/3}): W = m_drop · g = 2π · r · γ · φ(r / V^{1/3}), where V = m_drop / ρ is the single drop volume. By collecting and weighing detached droplets across six systematic drop counts (N = 10 to 60 drops) under strict thermostatic control at 20.0 °C with a detachment period of 8–10 s per drop, ordinary least-squares regression of cumulative mass against drop count extracts the unbiased single-drop mass m_drop as the slope. Inverting via Harkins-Jordan calibration isolates the absolute surface tension of pure water (72.80 mN/m) with sub-decimal precision, confirming the physics of interfacial thermodynamics.

m_total = m_drop · N + m₀ ==> m = 0.1004 · N

The linear regression models the modified Tate detachment equation m_total = [ (2π · r · γ · φ) / g ] · N + m₀, where N represents the integer count of detached droplets, m_total represents the cumulative gravimetric mass in g, and m₀ represents the initial weighing bottle tare offset in g (theoretically m₀ = 0.0000 g). Multiplying the regression slope m_slope by K_tip = g / (2π · r · φ) = 725.11 N/(m·kg) directly isolates the absolute surface tension in mN/m, while constant tare residues are absorbed into m₀.

5. Method

  1. Clean the stalagmometer with warm chromic-sulfuric acid cleaning mixture, rinse thoroughly with 500 mL of boiling deionized water, wash with absolute ethanol, and dry under filtered nitrogen to ensure the flat tip is perfectly hydrophilic and free of organic surfactant films.
  2. Mount the stalagmometer inside the circulating glass water jacket set to 20.0 °C. Charge the upper reservoir with ultra-pure deionized water and allow 15 minutes for thermal equilibrium.
  3. Adjust the micro-needle valve at the top of the stalagmometer until pendent droplets form smoothly and detach at a rate of exactly one drop every 8 to 10 seconds, ensuring quasi-static detachment without inertial jetting.
  4. Tare a clean, dry, stoppered weighing bottle on the analytical balance to ±0.0001 g. Position the bottle directly beneath the stalagmometer tip inside the draft-shielded enclosure.
  5. Uncap the weighing bottle, count exactly 10 drops as they detach from the tip into the bottle, immediately replace the stopper to prevent evaporation, and record the mass on the analytical balance.
  6. Repeat the procedure for 20, 30, 40, 50, and 60 drops. Regress total collected mass against drop count to obtain the single drop mass m_drop, compute the Harkins-Jordan factor φ(r/V^{1/3}), and calculate absolute surface tension γ.

6. Observations and calculation table

All gravimetric observations were recorded using a calibrated 4-decimal electronic analytical balance enclosed within a draft shield chamber. Volumetric aliquots were dispensed from a Class A burette. Residuals are computed relative to the ordinary least-squares regression line.

Table 1. Seven simulated gravimetric observations; x and y represent dispensed volume and net mass metrics.
No.Detached droplet count N (drops) (drops)Cumulative detached droplet mass m_total (g)Detached droplet count N (drops)Cumulative droplet mass m_total (g)Fitted y (g)Residual (g)
110.01.004010.00001.00401.00400.0000
220.02.008020.00002.00802.00800.0000
330.03.012030.00003.01203.01200.0000
440.04.016040.00004.01604.01600.0000
550.05.020050.00005.02005.02000.0000
660.06.024060.00006.02406.02400.0000

Download the sample observations (CSV) ↓

7. Worked calculation for one observation

Stalagmometer Capillary Tip Radius: r_ext = 3.500 mm = 0.00350 m; Fluid: Pure Deionized Water at 20.0 °C. Collected Droplet Count: N = 10 drops; Cumulative Droplet Mass: m_total = 1.0040 g. Single Droplet Gravimetric Mass: m_drop = m_total / N = 1.0040 g / 10 = 0.10040 g/drop. Single Droplet Volume: V = m_drop / ρ = 0.10040 g / 0.9982 g/cm³ = 0.10058 cm³ = 100.58 mm³. Dimensionless Shape Factor: r / V^(1/3) = 3.500 mm / (100.58 mm³)^(1/3) = 0.7526; Harkins-Jordan Factor φ = 0.6150. Single-Trial Dynamic Surface Tension: γ = (m_drop · g) / (2π · r · φ) = (0.000100 × 9.807) / (2π × 0.00350 × 0.6150) × 1000 = 72.80 mN/m. Verification: Across the full six-point series from 10 to 60 drops, ordinary least-squares regression extracts the single-drop mass slope m_slope = 0.10040 g/drop, scaling by K_tip = 725.11 N/(m·kg) to isolate the surface tension γ = 72.80 mN/m with R² = 1.00000, confirming Tate's law with Harkins-Jordan correction.

8. Graph, gradient and final calculation

0.34.01.919.53.535.05.150.56.866.0Detached droplet count N (drops)Cumulative droplet mass m_total (g)V = 10 mL, m = 1.004 gV = 20 mL, m = 2.008 gV = 30 mL, m = 3.012 gV = 40 mL, m = 4.016 gV = 50 mL, m = 5.02 gV = 60 mL, m = 6.024 g

Circular markers represent empirical gravimetric observations. The solid line represents the ordinary least-squares linear regression fit.

x̄ = 35.0000 mL; ȳ = 3.5140 g
Σ(x − x̄)² = 1750.0000 mL²
Σ(x − x̄)(y − ȳ) = 175.7000 g·mL
m = Σ(x − x̄)(y − ȳ) / Σ(x − x̄)² = 0.10040000 g/mL
b = ȳ − mx̄ = 0.00000000 g
γ = m_slope · K_tip = 0.100400 × 725.10 = 72.80 mN/m; Intercept m₀ = 0.000000 g

9. Uncertainty and comparison

Percentage difference = |estimate − theoretical value| / |theoretical value| × 100 = 0.0000%.

For n = 7 gravimetric observations, Σresidual² = 9.8608e-31; s² = Σresidual²/(n − 2). Gradient standard error = √[s²/Σ(x − x̄)²] = 1.1869e-17. Intercept standard error = √[s²(1/n + x̄²/Σ(x − x̄)²)] = 4.6222e-16.

Propagating the fit parameters yields an experimental density standard error of 8.6060e-15 mN/m, with an ideal coefficient of determination R² = 1.000000.

In drop-weight stalagmometry governed by Tate's modified law m_total = [ (2π · r · γ · φ) / g ] · N + m₀, analytical measurement uncertainty is partitioned into distinct mechanical, thermal, and interfacial components: (1) Stalagmometer Tip Radius Accuracy: The external ground-glass tip is measured using an optical measuring microscope to r = 3.500 ± 0.002 mm (δr / r = 0.057%). Because surface tension scales inversely with r (γ ∝ 1/r), this contributes a systematic geometric uncertainty of only ±0.04 mN/m. (2) Harkins-Jordan Factor Plateau Stability: At r / V^(1/3) = 0.7526, the Harkins-Jordan function φ(r/V^{1/3}) resides in a broad, nearly horizontal plateau where the derivative ∂φ/∂(r/V^{1/3}) ≈ 0. A 2% variation in droplet volume shifts φ by less than 0.08%, imparting extraordinary robustness against droplet sizing. (3) Thermostatic Jacket Control: Liquid water exhibits a negative temperature coefficient ∂γ/∂T ≈ -0.151 mN/(m·°C) at 20.0 °C. The circulating thermostatic water mantle maintains 20.0 ± 0.05 °C, restricting thermal surface tension drift to < ±0.008 mN/m (0.011%). (4) Evaporative Loss Prevention: Droplets detach directly into a narrow-mouth, ground-glass stoppered weighing bottle. Gravimetric evaporation trials verify that vapor loss is less than 0.0002 g over a 10-minute collection interval (<0.005% of 4.0 g). (5) Quasi-Static Detachment Kinetics: Throttling flow to 1 drop every 8 to 10 seconds suppresses hydrodynamic inertial forces (We = ρ v² r / γ < 10⁻⁴), ensuring pure capillary pinch-off without jetting. Ordinary least-squares linear regression across six independent drop counts (10 to 60 drops) yields an empirical slope of m_slope = 0.10040 g/drop with an ideal coefficient of determination R² = 1.00000, isolating the absolute surface tension γ = 72.80 mN/m with zero residual offset.

10. Result and discussion

SurfaceTension_gamma = 72.80 mN/m
CIPM standard reference value: 72.80 mN/m

The experimental observations provide rigorous quantitative verification of Interfacial Thermodynamics, Tate's Law of Capillary Detachment, and the Harkins-Jordan Hydrodynamic Correction. In liquid interfacial physics, surface tension arises from the unbalanced cohesive intermolecular forces pulling surface molecules into the bulk liquid phase. During the growth of a pendent drop at the razor-flat circular stalagmometer tip, the downward gravitational weight increases until it overcomes the upward surface tension force. Crucially, high-speed fluid mechanics dictates that when the capillary neck pinches off via Rayleigh-Plateau instability, approximately 38.5% of the droplet volume remains behind on the glass tip. Incorporating the Harkins-Jordan factor φ = 0.6150 (derived for the dimensionless shape ratio r / V^{1/3} = 0.7526) completely compensates for this incomplete detachment. Across all six drop count determinations from 10 to 60 drops, cumulative mass scaled in strictly linear proportion to drop count. The ordinary least-squares regression line m_total = 0.10040 · N + 0.0000 g yielded an ideal coefficient of determination R² = 1.00000. Scaling by K_tip extracted an absolute surface tension of γ = 72.80 mN/m, in perfect agreement with the international reference standard for pure water at 20.0 °C. Throttling the flow to 8–10 seconds per drop ensured quasi-static detachment, and acid cleaning eliminated trace surfactant films that would otherwise depress surface tension.

11. Precautions and limitations

The determination of surface tension via drop-weight stalagmometry is subject to specific experimental constraints: (1) Capillary Edge Boundary Pinning: Wetting fluids must pin cleanly to the sharp external circumference of the ground-glass tip; if the glass is contaminated, the liquid film can creep up the outer cylindrical wall, increasing the effective radius r. (2) Dynamic Surface Age and Adsorption Kinetics: For pure water, surface tension equilibrates in milliseconds; however, for surfactant solutions, slow diffusion to the dilating interface produces dynamic surface tension effects requiring long bubble-pressure or drop-shape analysis. (3) Satellite Droplet Recoil: During Rayleigh-Plateau neck rupture, microscopic satellite droplets (<1% of mass) can recoil or mist into ambient air. (4) Evaporative Vapor Deficit: Over a 60-drop run lasting ~10 minutes, minute evaporation through the needle-valve vent must be controlled with vapor-saturated pre-chambers. The drop-weight stalagmometry physical model embodies the relation m_total = m_drop · N + m₀, where m_drop = (2π · r · γ · φ) / g. Calibrated apparatus constants: stalagmometer external tip radius r = 0.003500 m (3.500 mm), pure water surface tension γ = 0.07280 N/m (72.80 mN/m) at 20.0 °C, water density ρ = 998.2 kg/m³, standard acceleration of gravity g = 9.80665 m/s². The single-drop volume is V = 0.10058 cm³ = 1.0058 × 10⁻⁷ m³, giving the dimensionless shape parameter r / V^{1/3} = 0.00350 / (1.0058 × 10⁻⁷)^{1/3} = 0.7526. The corresponding Harkins-Jordan correction factor is exactly φ = 0.6150. Under these parameters, the theoretical single-drop mass is m_drop = (2π · 0.00350 · 0.07280 · 0.6150) / 9.80665 = 0.00010040 kg = 0.10040 g/drop. Multi-point linear regression across N = 10 to 60 drops yields an empirical slope of m_slope = 0.10040 g/drop with R² = 1.00000. Any constant weighing bottle tare offset or initial wetting residue is absorbed into m₀ = 0.0000 g, guaranteeing that the slope serves as an unbiased estimator of single-drop mass.

12. Conclusion

This laboratory investigation successfully determined the absolute liquid surface tension of ultra-pure water at 20.0 °C using precision drop-weight stalagmometry. Through systematic multi-drop gravimetric determinations across counts from N = 10 to 60, we extracted an absolute surface tension of γ = 72.80 mN/m with a fit standard error of 8.6060e-15 mN/m and an ideal coefficient of determination R² = 1.00000. The observed surface tension confirms the IAPWS thermodynamic reference value (72.80 mN/m) with a percentage difference of 0.0000%. The investigation verified Tate's law, proved the necessity of the Harkins-Jordan hydrodynamic detachment correction, and demonstrated that multi-point gravimetric regression insulates interfacial measurements from container taring and transfer biases.