EXPERIMENT 47 · PHYSICS LABORATORY
Internal Resistance and Electromotive Force of a Cell: Terminal Voltage and Load Characteristics
To determine the electromotive force (EMF) and internal resistance of an electrochemical primary cell by measuring terminal potential difference as a function of load current, evaluate the short-circuit current and maximum power transfer, and investigate instrument loading effects. Explore a virtual apparatus, collect readings and analyse a graph.
FOLLOW THE METHOD
A reason behind every step.
To determine the electromotive force (EMF) and internal resistance of an electrochemical primary cell by measuring terminal potential difference as a function of load current, evaluate the short-circuit current and maximum power transfer, and investigate instrument loading effects. Suggested independent settings: 50.0, 100.0, 150.0, 200.0, 250.0, 300.0 mA.
1Open-Circuit Inspection and Baseline EMF Verification
Inspect the open-circuit cell using the high-impedance digital voltmeter with the main knife switch open; record the quiescent open-circuit terminal voltage to verify baseline electromotive force ℰ₀ = 1.500 V prior to drawing discharge current.
Why this step matters
Establishing the open-circuit EMF before current flow verifies the chemical integrity of the cell and confirms that zero initial polarization exists.
2Closed-Loop Series Circuit and Kelvin Voltmeter Connections
Connect the circuit in series: primary cell positive terminal to ammeter input, ammeter output to the decade resistance box, and resistance box back to the cell negative terminal through the knife switch. Connect the high-impedance voltmeter directly across the cell battery posts (Kelvin voltage sensing configuration) to exclude lead resistance drops.
Why this step matters
Kelvin 4-wire potential sensing ensures that the voltmeter measures the true terminal voltage across the cell electrodes, completely bypassing ohmic drops along the high-current connecting wires.
3Initial High-Resistance Load Setting and Current Reading
Set the decade resistance box to its highest load setting (producing minimum initial current I = 50.0 mA). Momentarily close the knife switch for 3 to 5 seconds, read both the ammeter current I and the digital voltmeter terminal potential difference V, and immediately open the switch to prevent chemical exhaustion or polarization buildup.
Why this step matters
Starting at minimum current limits initial chemical consumption, while brief 3 to 5 second pulses prevent cumulative concentration polarization and internal Joule heating.
4Chemical Relaxation Window and Stepwise Load Adjustment
Allow the cell a 60-second recovery period for ionic diffusion to re-establish chemical equilibrium at the electrode interfaces. Step down the load resistance progressively to draw currents of 100.0 mA, 150.0 mA, 200.0 mA, 250.0 mA, and 300.0 mA, recording the stabilized terminal voltage V at each setting.
Why this step matters
The 60-second recovery interval allows depolarizing agents to clear accumulated hydrogen films, ensuring each observation begins from identical equilibrium conditions.
5Data Tabulation and Linear Regression Fitting
Tabulate the paired values of load current I (in amperes) and terminal potential difference V (in volts). Plot terminal voltage V on the vertical axis against current I on the horizontal axis; fit an ordinary least-squares straight line to determine the slope m and the vertical intercept b.
Why this step matters
Linear regression across multiple discharge levels averages out meter quantization noise and contact resistance fluctuations, yielding statistically robust fit parameters.
6Parameter Extraction and Performance Bounds Calculation
Extract the internal resistance as r = −m and the true electromotive force as ℰ = b. Calculate the extrapolated short-circuit current I_sc = ℰ / r, evaluate the maximum deliverable external power P_max = ℰ² / (4r), and compare the fitted parameters against manufacturer reference specifications.
Why this step matters
Extrapolating from the linear regression determines the absolute physical operating bounds of the cell without requiring destructive short-circuit testing.