EXPERIMENT 25 · PHYSICS LABORATORY
Verification of Newton's Law of Cooling and Determination of the Cooling Constant
Investigate the rate of loss of heat by radiation and convection for a body cooling in a constant-temperature enclosure, verifying Newton's law of cooling for small temperature differences, and determining the cooling constant k from the linear logarithmic temperature excess decay. Explore a virtual apparatus, collect readings and analyse a graph.
FOLLOW THE METHOD
A reason behind every step.
Investigate the rate of loss of heat by radiation and convection for a body cooling in a constant-temperature enclosure, verifying Newton's law of cooling for small temperature differences, and determining the cooling constant k from the linear logarithmic temperature excess decay. Suggested independent settings: 2.0, 4.0, 6.0, 8.0, 10.0 min.
1Jacket Temperature Stabilization and Baseline Verification
Fill the annular space of the double-walled enclosure with room-temperature water. Stir the jacket bath thoroughly and record its steady temperature T_0 using a reference thermometer. Confirm that T_0 remains stationary at 20.0 °C throughout the duration of the laboratory session, providing a constant ambient thermal sink.
Why this step matters
Newton's law of cooling assumes an invariant environmental boundary temperature T_0. Any unmodelled drift in the enclosure temperature directly alters the instantaneous driving potential (T − T_0), distorting the apparent exponential decay rate.
2Calorimeter Preparation, Water Tare, and Heating Charge
Weigh the clean, dry polished copper calorimeter vessel with its stirrer to ±0.01 g. Add exactly 100.0 g of distilled water preheated in a separate beaker to approximately 85 °C. Mount the calorimeter securely inside the central chamber of the water-jacketed enclosure, seating it firmly upon insulating vulcanized cork pegs to eliminate conductive thermal bridging to the bottom plate.
Why this step matters
Heating the water prior to introducing it into the vessel ensures that the calorimeter metal and water rapidly reach thermal equilibrium near 80 °C without exposing the enclosure jacket to open flames or heating coils.
3Lid Closure, Thermometer Immersion, and Time Zero Synchronization
Place the varnished wooden lid firmly over the calorimeter, insert the calorimetric thermometer through the central rubber grommet so its bulb is submerged in the geometric center of the liquid, and insert the loop stirrer. Stir the hot water gently until the thermometer stabilizes at the target initial temperature T_i = 80.00 °C. Start the digital bench chronometer simultaneously as time t = 0.0 min.
Why this step matters
Central bulb placement avoids boundary-layer edge effects near the metallic walls, ensuring the thermometer samples the bulk core liquid. Synchronizing timer start with a precise initial temperature sets the exact boundary condition θ_0 = 60.00 °C.
4Periodic Agitation and Serial Temperature Acquisition
Maintain continuous, gentle vertical agitation with the loop stirrer at approximately one stroke every two seconds to prevent vertical thermal stratification. At exact half-minute intervals, inspect the thermometer meniscus using an optical magnifier to eliminate parallax, recording the water temperature T at 2.0, 4.0, 6.0, 8.0, and 10.0 minutes of elapsed time.
Why this step matters
Because water has a low thermal conductivity (k ≈ 0.6 W m⁻¹ K⁻¹), thermal stratification occurs rapidly in stagnant cooling liquid, creating cold upper layers and hot lower pools. Gentle stirring maintains spatial isothermality without inducing viscous friction heating.
5Logarithmic Temperature Excess Transformation
For each recorded observation time t, compute the instantaneous temperature excess θ(t) = T(t) − T_0 above the jacket temperature. Evaluate the logarithmic decay ratio y = ln[θ_0 / θ(t)] = ln[60.00 / (T(t) − 20.00)], where θ_0 = 60.00 °C is the initial temperature excess at t = 0.
Why this step matters
Transforming the exponential cooling trajectory into natural logarithmic coordinates linearizes the underlying differential equation, mapping elapsed time to a straight line whose slope directly represents the Newton cooling constant k.
6Linear Least-Squares Regression and Parameter Estimation
Plot the logarithmic temperature excess y against elapsed cooling time t (min). Perform an ordinary least-squares linear regression to extract the slope m, standard error, and coefficient of determination R². Confirm that the gradient m equals the system cooling constant k = 0.0960 min⁻¹, and compute the thermal time constant τ = 1/k.
Why this step matters
Linear regression over multiple serial points distributes instrument reading uncertainty across the full cooling arc, suppressing random parallax errors and rigorously testing the validity of the first-order cooling hypothesis through the goodness-of-fit statistic R².