EXPERIMENT 52 · PHYSICS LABORATORY
Photometry and Lamp Efficiency
Verify the inverse-square law of illumination E = I / r², determine the luminous intensity I of an electric lamp in candelas (cd) using a precision optical bench photometer, calculate total luminous flux Φ = 4π I in lumens, and evaluate the luminous efficacy η = Φ / P in lumens per watt across incandescent, compact fluorescent (CFL), and solid-state LED light sources. Explore a virtual apparatus, collect readings and analyse a graph.
FOLLOW THE METHOD
A reason behind every step.
Verify the inverse-square law of illumination E = I / r², determine the luminous intensity I of an electric lamp in candelas (cd) using a precision optical bench photometer, calculate total luminous flux Φ = 4π I in lumens, and evaluate the luminous efficacy η = Φ / P in lumens per watt across incandescent, compact fluorescent (CFL), and solid-state LED light sources. Suggested independent settings: 0.6, 0.8, 1.0, 1.25, 1.5, 1.75, 2.0 m.
1Bench Leveling and Optical Axis Collimation
Mount the optical bench on a vibration-isolated laboratory table and adjust the leveling screws until the spirit level indicates zero tilt along both longitudinal and transverse axes. Install the test lamp housing, photometer contrast head, and sensor carrier on their sliding riders. Adjust the vertical and lateral spindle thumbscrews until the centers of the lamp filament, photometer diffusion screen, and detector aperture are collinear with the central optical axis of the track.
Why this step matters
Any lateral or vertical misalignment causes light rays to strike the photometer screen at oblique angles rather than normal incidence. By Lambert's cosine law, oblique illumination reduces irradiance by cos θ, introducing severe systematic errors into the inverse-square distance calculations.
2Standardization and Dark-Current Zeroing
Close the light-tight shielding curtains around the optical bench to eliminate ambient room lighting. Energize the digital luxmeter and power analyzer without lighting the lamps, and verify that the residual dark-current display registers zero lux (tare zero adjustment). Light the primary reference standard lamp and allow it to operate for 10 minutes until thermal and filament resistance equilibrium are achieved.
Why this step matters
Stray ambient illumination from room windows or ceiling lamps adds directly to the measured illuminance, destroying the strict inverse-square relation E ∝ 1/r². Thermal stabilization ensures that filament resistance and color temperature remain perfectly constant throughout all subsequent trials.
3Mounting and Electrical Calibration of the Test Lamp
Mount the test incandescent lamp (75 W nominal rating) into the test socket and connect its terminals to the regulated power analyzer. Adjust the power supply voltage until the operating voltage reaches exactly 120.0 V RMS. Record the active power consumption P (in Watts), current I_A, and verify that the electrical input remains steady within ±0.05 W.
Why this step matters
Tungsten incandescent filament luminous intensity is extraordinarily sensitive to operating voltage, scaling approximately as I_lum ∝ V^3.5. Maintaining exact electrical power regulation is indispensable for obtaining reproducible luminous efficacy measurements.
4Systematic Distance Variation and Illuminance Measurement
Position the photometer detector at an initial separation distance r = 0.60 m from the effective center of the lamp filament. Verify that the detector face is perpendicular to the beam axis. Record the steady-state illuminance E displayed on the calibrated luxmeter. Systematically increment the distance to r = 0.80 m, 1.00 m, 1.25 m, 1.50 m, 1.75 m, and 2.00 m, recording the steady illuminance at each station.
Why this step matters
Sampling seven widely spaced distances spanning a four-fold distance range (0.5 m to 2.0 m) produces a sixteen-fold range in illuminance, providing an expansive dynamic range for linear regression of E against 1/r² with high statistical power.
5Visual Null Comparison with Lummer-Brodhun Head
Replace the photoelectric sensor with the Lummer-Brodhun contrast photometer head, placing the certified standard lamp (I_s = 25.0 cd) at the opposite end of the 2.00 m track. Slide the photometer carriage along the bench until the concentric circular fields viewed through the eyepiece exhibit identical brightness and zero contrast boundary. Record the balancing distances r_x and r_s.
Why this step matters
Visual null-detection provides an independent optical verification of the luminous intensity via the classic equality I_x / r_x² = I_s / r_s², corroborating electronic luxmeter readings with direct human psychophysical photometric metrology.
6Interchange of Light Sources and Comparative Efficacy Audit
Replace the 75 W incandescent lamp sequentially with a 14 W compact fluorescent lamp (CFL) and a 9 W solid-state LED lamp designed for equivalent illumination. For each source, record the steady-state electrical power P, determine the luminous intensity I at r = 1.00 m, calculate total luminous flux Φ = 4π I, and compute the luminous efficacy η = Φ / P in lumens per watt.
Why this step matters
Comparing different lamp technologies under identical geometric and photometric bench conditions demonstrates the dramatic evolutionary leap in lighting efficacy from incandescent blackbody radiation (~16 lm/W) to phosphorescent gas discharge (~70 lm/W) and solid-state bandgap electroluminescence (>110 lm/W).