Physics / Potentiometer EMF Comparison
FREE TO EXPLORE
EXPERIMENT 51 · PHYSICS LABORATORY

Potentiometer EMF Comparison lab report

A worked potentiometer emf comparison lab report with simulated observation tables, graph, calculations, uncertainty and results.

WORKED EXAMPLE · SIMULATED OBSERVATIONS

A complete example report.

Simulated demonstration data.

This report shows the method and arithmetic. It is not evidence of a physical experiment. Replace its observations with your own measurements when preparing a real lab report.

Student: __________________Student ID: __________________Course: __________________Date: __________________Partner: __________________Instructor: __________________

1. Title and aim

The Potentiometer: Accurate Measurement and Comparison of Electromotive Forces. To calibrate a slide-wire potentiometer using a Weston standard cell, determine the uniform potential gradient along the slide wire, compare the electromotive forces of dry and primary cells at null balance without drawing current, and verify that cell EMF is strictly proportional to balancing length.

2. Abstract

Five observations were generated from the ideal teaching model over the settings listed below. A straight-line fit gave an estimate of potentiometer potential gradient of 2.0000 V m⁻¹, differing by 0.000% from the model value. The reported scatter comes from scale rounding and any stated model approximation, not measured laboratory noise.

3. Apparatus and fixed settings

4. Theory and graph choice

The slide-wire potentiometer, invented by Johann Christian Poggendorff in 1841, is the fundamental null-balance instrument for the precision measurement and comparison of electromotive forces. Unlike deflection voltmeters that draw operating current and depress terminal voltage via internal resistance drops (V = ℰ − I r), the potentiometer opposes the unknown cell EMF against a known, continuous potential drop along a uniform resistance wire. When the sliding contact is adjusted until the sensitive center-zero galvanometer indicates zero current (I_G = 0), the wire potential drop exactly equals the open-circuit cell EMF without perturbing the cell's thermodynamic equilibrium. In this experiment, students calibrate the wire potential gradient (k = 2.000 V/m = 0.02000 V/cm) using a saturated Weston cadmium standard cell (ℰ_s = 1.0186 V at ℓ_s = 50.93 cm), measure balancing lengths across six distinct cell EMF levels spanning 0.400 V to 1.500 V, apply linear regression to determine k and verify Poggendorff’s law (ℰ_x / ℰ_s = ℓ_x / ℓ_s), and explore the zero-current infinite input impedance principle.

ℰ = k · L + b; k = m_slope, ℰ_x / ℰ_s = ℓ_x / ℓ_s

The calculation fits y = mx + b in SI units. The graph may display scaled axes for readability; those display factors are removed before calculating the physical result. The gradient has units V m⁻¹.

5. Method

  1. Connect the regulated 2.000 V driver accumulator in series with a master knife switch, an adjustable wirewound rheostat, and the 1-meter potentiometer slide wire AB. Ensure the positive terminal of the driver accumulator connects directly to binding post A (the zero of the metric scale). Close the switch and adjust the rheostat to set the potential drop across wire AB to slightly greater than the maximum cell EMF to be measured (setting the potential gradient to exactly k = 0.02000 V/cm).
  2. Connect the positive terminal of the Weston cadmium standard cell (ℰ_s = 1.0186 V) to the common positive post A. Connect the negative terminal through the protective 10,000 Ω resistor and sensitive center-zero galvanometer to the movable knife-edge jockey slider F. Verify that the standard cell electromotive force directly opposes the potential drop along the slide wire from post A.
  3. Lightly touch the jockey contact to the extreme left of the wire near post A (ℓ ≈ 1 cm) and tap the galvanometer key, observing the direction of needle deflection. Next, touch the jockey to the extreme right near post B (ℓ ≈ 99 cm) and tap the key. Verify that the two deflections occur in opposite directions. If opposite deflections are not obtained, reverse the cell connections or reduce the driver rheostat resistance.
  4. Use the method of successive approximations (bisection search) to locate the approximate null point with the 10,000 Ω protective resistor in series. Once the needle deflection is less than 1 scale division, depress the bypass key to short-circuit the protective resistor, and gently adjust the jockey position until the galvanometer rests at perfect zero. Record the standard cell balance length ℓ_s = 50.93 cm and compute the potentiometer potential gradient k = ℰ_s / ℓ_s = 1.0186 / 50.93 = 0.02000 V/cm (2.000 V/m).
  5. Switch the secondary branch to the unknown test cells using the selector switch, ensuring that the driver circuit rheostat remains completely untouched. For each of the six cell EMF levels (spanning 0.400 V to 1.500 V, including standard cell, Daniell cell, and dry cells), determine the precise balancing length ℓ_x at galvanometer null. Record the balancing length in centimeters and convert to meters (L_x = ℓ_x / 100.0).
  6. Plot the balancing length L (in meters) on the horizontal axis and the cell electromotive force ℰ (in Volts) on the vertical axis. Compute ordinary least-squares linear regression to determine the potential gradient k (the slope m in V/m) and the vertical intercept b. Calculate the cell EMF ratios ℰ_x / ℰ_s and compare against the corresponding length ratios ℓ_x / ℓ_s to verify Poggendorff’s law of the potentiometer.

6. Observations and calculation table

All observations use fixed model settings. Values are rounded only at the instrument reading stage; further calculations use full precision. Repeated physical trials are not claimed.

Table 1. Five simulated measurements; x and y are in SI units.
No.Test cell electromotive force ℰ (V)Null balance length ℓ (cm)x (SI)y (SI)Fitted y (SI)Residual (SI)
10.400020.000.200000.400000.400000.00000
20.650032.500.325000.650000.650000.00000
30.900045.000.450000.900000.900000.00000
41.018650.930.509301.018601.018600.00000
51.250062.500.625001.250001.250000.00000
61.500075.000.750001.500001.500000.00000

Download the sample observations (CSV) ↓

7. Worked calculation for one observation

Cell electromotive force ℰ = 0.400 V; observed null balancing length ℓ = 20.00 cm (L = 0.2000 m); potential gradient estimate k_single = ℰ / L = 0.400 / 0.2000 = 2.0000 V m⁻¹ (0.02000 V cm⁻¹). Linear regression of ℰ against L across all five cells yields the reported slope k and intercept b.

8. Graph, gradient and final calculation

0.290.150.620.310.950.481.280.641.610.81Balancing length L (m)Electromotive force ℰ (V)

Dots show the sample observations. The line is the ordinary least-squares fit.

x̄ = 0.47655; ȳ = 0.95310
Σ(x − x̄)² = 0.19804
Σ(x − x̄)(y − ȳ) = 0.39607
m = Σ(x − x̄)(y − ȳ) / Σ(x − x̄)² = 2.00000000
b = ȳ − mx̄ = 0.00000000
Slope S = k = 2.0000 V m⁻¹ (0.02000 V cm⁻¹); Intercept b = 0.0000 V; Potential gradient k = 2.0000 V m⁻¹ (reference: 2.0000 V m⁻¹); Standard cell balance length ℓ_s = ℰ_s / k = 1.0186 / 0.02000 = 50.93 cm; Ratio law verified: ℰ_x / ℰ_s = ℓ_x / ℓ_s

9. Uncertainty and comparison

Percentage difference = |estimate − model value| / |model value| × 100 = 0.0000%. Percentage difference is not experimental uncertainty.

For n = 5 observations, Σresidual² = 0.00000; s² = Σresidual²/(n − 2). Gradient standard error = √[s²/Σ(x − x̄)²] = 0.00000. Intercept standard error = √[s²(1/n + x̄²/Σ(x − x̄)²)] = 0.00000.

Propagating only the relevant fit coefficient gives a fit-only standard error of 0.00000 V m⁻¹. This is a one-standard-error quantity under the regression assumptions. Deterministic rounded model data do not supply a complete physical uncertainty budget.

Slide-wire cross-sectional non-uniformity and local mechanical wear from knife-edge jockey contact (~0.1% diameter variation producing subtle local potential gradient fluctuations), primary driver accumulator voltage drift and internal heating during prolonged continuous current flow (minimized by using a heavy-capacity lead-acid accumulator and monitoring standard cell balance ℓ_s periodically), parasitic thermoelectric Seebeck EMFs at the bimetallic copper-constantan binding post junctions (up to ±15 μV if terminal posts are touched by warm fingers, cancelled by current reversal), galvanometer pivot friction and visual crosshair null detection resolution (±0.2 mm along the metric scale, corresponding to ±40 μV in potential), knife-edge slider contact resistance fluctuations, and connecting lead resistance between the zero index and the binding post govern the uncertainty envelope. Ordinary least-squares regression of cell electromotive force ℰ against balancing length L averages out random slider visual alignment errors and local wire imperfections, with the slope isolating the pure potential gradient k and the vertical intercept isolating any end-resistance or zero-index offset.

10. Result and discussion

k = 2.0000 V m⁻¹ (0.02000 V cm⁻¹) = 2.0000 V m⁻¹
Model comparison value: 2.0000 V m⁻¹

The observations follow the stated model relationship. Check whether the intercept and residual pattern suggest a systematic offset or a missing approximation. Agreement with the model does not independently verify its input constants.

11. Precautions and limitations

Calculations assume a perfectly uniform wire cross-section and resistivity, an unvarying driver accumulator current throughout all runs, negligible thermoelectric EMFs at metal junctions, and negligible contact resistance at the jockey slider. Real slide wires exhibit minor local diameter variations (~0.1%) caused by manufacturing tolerances or mechanical wear from slider friction, as well as parasitic thermal Seebeck EMFs at the copper-constantan binding posts if touched by warm fingers.

12. Conclusion

The worked example illustrates how to obtain potentiometer potential gradient from measurements, graph variables and a fitted coefficient. The estimate is 2.0000 V m⁻¹. A real experiment must additionally evaluate the apparatus-specific uncertainties and corrections described above.